Model 2000TM
Controller
Any QCM, Any Combination,
Up to 12 Channels

The Model 2000 Control/Data
Acquisition Unit (M2000) is the very best
lab-quality controller on the market for
control and data gathering functions
associated with quartz crystal
microbalances.
The Model 2000 was designed to
meet the growing need for a
cost-effective, yet efficient means of
interfacing any combination of up to
twelve of our various QCM mass sensors
(MK9, MK10, MK16, MK17, MK18, MK19, MK21,
MK23 and MK24) to a general purpose host
computer.
With a minimal front panel, the M2000
communicates with the host computer over
an industry standard RS232 serial
link. It accommodates both Si-diode
and PRT temperature sensors, and will
drive wire wound heaters, PRT heaters,
and single, dual and 3-stage
thermoelectric elements.
Additionally, it is able to power an
external heater, such as is used for an
effusion cell.
While autonomously controlling the
temperature, both temperature and
frequency data for each channel are sent
to the host over the serial link.
The base unit can handle four channels,
expandable to twelve channels in steps of
four. The data is sent to the host
at a rate of one reading per second per
channel.
The temperature data sent to the host
is given in degrees Kelvin, with the
least significant digit representing
1/100 of a Kelvin (0.01 K). The
electronics add no more than 1/4 of a
Kelvin (<0.25 K) error to the accuracy
of the sensor for absolute
accuracy. For relative accuracy,
which eliminates thermal drift, the added
error is less than 1/10 of a Kelvin
(<0.1 K).
Frequency data sent to the host is
given in Hertz. The relative
accuracy is better than one part per
million (>1 ppm). For all
frequencies (1 KHz to 600 KHz), seven
digits with a floating decimal point is
sent. Thus, for a frequency of 100
KHz, the least significant digit
represents 1/10 Hz (0.1 Hz).
The
software
included is simple to install and use,
with an integral programming language
which allows for simple command
sequences, if desired. A host
computer with software installed can be
supplied, if requested, or an in-house
computer may be used.


| Specifications |
| Sensors |
Si-Diode,
PRT |
Scale |
Hertz |
| Scale |
Kelvin |
Accuracy
(Relative) |
>1 ppm |
| Resolution |
0.01 K |
Range |
1 to 600 KHz |
Accuracy
(Absolute)
Accuracy (Relative) |
>0.25 K
>0.1 K |
Resolution |
7 Digits Floating
point, LSD for 100KHz is 0.1Hz |
| Weight |
16.5 lbs. (7.5 kg) |
Operating
Temp. |
10 to 40șC |
| Power
Requirement |
300 Watts,
(with 12 TQCMs)
90-264 VAC,
47-440 Hz |
Dimensions |
19" x
8.75" x 13"
48.25cm x 22.22cm x 56.45cm |
|